American Standard Medical Gas Manifold Air With Temperature Monitor
The automatic manifold is designed for automatic switching between tow gas sources. This series of automatic manifold can be automatically switched without any adjustment by the operator, even if the electrical system is powered off.
Specifications
Gas Service | Inlet Pressure | Outlet Flow | Outlet Pressure | Outlet Connection | Inlet Connection | Power |
O2 | 20Mpa | 100(3530)m3/h | 0.55Mpa | 3/4" loos joint(screw thread:RC 3/4) | Brass tube | 220V 50Hz |
Air | 20Mpa | 100(3530)m3/h | 0.8Mpa | 3/4" loos joint(screw thread:RC 3/4) | Brass tube | 220V 50Hz |
N2O | 20Mpa | 50(1765)m3/h | 0.5Mpa | 3/4" loos joint(screw thread:RC 3/4) | Brass tube | 220V 50Hz |
CO2 | 20Mpa | 30(1060)m3/h | 0.8Mpa | 3/4" loos joint(screw thread:RC 3/4) | Brass tube | 220V 50Hz |
Inert | 20Mpa | 100(3530)m3/h | 0.8Mpa | 3/4" loos joint(screw thread:RC 3/4) | Brass tube | 220V 50Hz |
Installation Dimension
Gas Service | W | H1 | H2 |
O2, Air, N2O, Inert gas | 540mm | 350mm | 600mm |
CO2 | 650mm | 350mm | 600mm |
Features
About us
Kingbell International Technology Development Co., Ltd. is a professional high-tech enterprise specialized in clean operation room system and central gas supply system. We register in Hong Kong and with factory in Hua Xin town of shanghai, China.
We are focused on construction of clean OTs, clean industrial plants (pharmaceutical factory, food factories, electronics factory, cosmetics factory etc.) and clean laboratories. We have professional teams including designing, budgeting, constructing, commissioning and testing, which can ensure the quality of projects. And our team members have participated in more than 100 hospital and clean workshop projects, has accumulated rich experience in this field.